FloormapX System Overview
30
4.
Charging is complete once a green LED is displayed on the battery charger. If the LED indicates
a different color, please refer to the table below, or the information on the charger to identify
the mode.
5.
Switch off the battery charger power at the mains supply, then disconnect from the mains
supply.
IMPORTANT
Nickel Metal hydride batteries require a regular charge and discharge cycle to maintain
optimal performance. It is important that the battery is allowed to be fully charged between
discharge cycles during this period. Failure to do so will invalidate the warranty.
Battery Charger LED Indications
The color of the LED indicator on the charger correlates to the charging mode, which can be identified
from the table below:
LED Color
Mode
Yellow
Battery not connected
Yellow
Battery initialization & analysis
Orange
Fast charge
Green with intermittent Yellow flash
Top-off charge
Green
Trickle charge
Alternating Orange - Green
Error
Battery Safety Information
•
The FloormapX batteries should be charged at least every 4 months, otherwise the batteries
will reduce capacity or become dead.
•
The FloormapX batteries must be fully charged before use. Allow 3-5 cycles of charging and
discharging for full battery capacity to be recovered.
•
The FloormapX battery charger is only designed for indoor use and should not encounter water
or dust. To avoid overheating, the charger and battery should not be covered when it is in use.
•
When the FloormapX system is not being used the battery must be removed from the right-
hand active battery slot of the scanner and placed into the left-hand storage slot. The battery
capacity has a potential of discharging if left in the active slot when the system is not being
used.
•
Only use the charger provided to charge the FloormapX battery. Charging the battery with
another charger or without any attention may cause the battery to explode.
•
The charger is turned on by connecting it to the mains power socket. Disconnecting it from the
main power socket turns the charger off. The mains power socket should be easily accessible. If
an operational error occurs, the plug should be immediately removed from the socket.
•
If the charger is equipped with a mains cord, verify that the cord has not been damaged. If the
cord is damaged, the charger must not be used.
•
The charger contains dangerous voltages, so the cover should not be removed. Avoid the
plastic casing encountering any chemicals or solvents such as oil, greases, etc. as most types
of plastic can be broken down then. If the casing or covers are damaged, the charger must not
be used.
Содержание Floormap X
Страница 1: ...User s Manual Floormap X MFL Array Tank Floor Inspection Solution...
Страница 10: ...10 Chapter 1 Precautions and conventions...
Страница 16: ...General Precautions and Conventions 16 1 5 EU Declaration of Conformity...
Страница 19: ...19 Chapter 2 FloormapX System Overview...
Страница 35: ...35 Chapter 3 Inspection Preparation Setup and Operation...
Страница 43: ...43 Chapter 4 SIMS GO Data Acquisition Software...
Страница 69: ...69 Chapter 5 SIMS PRO...
Страница 98: ...98 Chapter 6 Troubleshooting and Maintenance...