Adjustment range for default alarm limits:
Fixed alarm limits / times:
Apnea pressure
after 15 seconds
Apnea flow
after 15 seconds
Apnea CO
2
after 15 seconds (after 60 seconds
in MAN/ SPONT)
– – : The alarm limit is deactivated as factory default.
NOTE: The newly set default alarm limits will be in
effect whenever the system is switched on or started
up from standby mode.
Certain alarm limits are switched off automatically in
MAN/SPONT: see tables, alarm limits for
MAN/SPONT, page 49.
Alarm limits
Adjustment range
Factory default
SpO
2
[%]
_
51 to 100
50 to 99
– –
92
Pulse
ª
[1/min]
_
31 to 300
30 to 299
120
50
EtCO
2
[mmHg]
_
1 to 75
0 to 74
50
– –
FiCO
2
[mmHg]
_
0 to 10
– –
5
– –
MV
[L/min]
_
0.1 to 39
0 to 38.9
12
3.0
FiO
2
[Vol.%]
_
19 to 100
18 to 99
– –
20
Fi Hal.
[Vol.%]
_
0.1 to 7
0 to 6.9
1.5
– –
Fi Iso.
[Vol.%]
_
0.1 to 7
0 to 6.9
2.3
– –
Fi Enf.
[Vol.%]
_
0.1 to 7
0 to 6.9
3.4
– –
Fi Des.
[Vol.%]
_
0.1 to 21.9
0 to 21.8
12.0
– –
Fi Sev.
[Vol.%]
_
0.1 to 9.9
0 to 9.8
3.4
– –
PAW
[cmH
2
O]
_
5 to 98
0 to 35
40
8
Configuring in Standby Mode
Setting Default Values
Operating Instructions Julian SW 3.n
97
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...