![Dräger Julian Скачать руководство пользователя страница 58](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787058.webp)
Leak Test
– When the soda lime has been changed
or
– when the breathing hoses or breathing system has
been changed.
– if the Vapor has been changed or topped up
●
test with the handwheel set to »0« and
●
test with the handwheel set to less than 0.2% by vol.
●
After the leak test, turn handwheel to »0«.
1 Close the Y-piece = place firmly on the cone.
2 Ensure that the sample line is not connected to the
patient circuit.
3 If necessary, seal the sample-line connector with
a luer-lock cap.
●
Press the softkey »leakage« in standby mode.
Julian performs the leak test for CMV/PCV and
determines the volume correction for system
compliance, duration approx. 30 seconds.
The breathing bag and its hose are not included in the
test.
Display:
When the test has been completed successfully, the
clock symbol disappears and the values for leakage
and system compliance are displayed.
The results of the leak test are also displayed on the
data screen page.
To return to the Standby screen:
●
Press dial knob.
WARNING !
Do not perform leak test while a patient is
connected to the anesthesia workstation.
Operation
Leak Test
Operating Instructions Julian SW 3.n
58
2
1
3
Leak Test
System
Compliance
Leak
cancel
test
Please wait
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...