Ventilating Children
For tidal volumes V
T
of less than 200 mL:
●
Use pediatric breathing circuit
Connecting the breathing circuit
1 Connect 0.5 L breathing bag with connector to the
breathing hose with the large connection sleeves.
Slip breathing hose over the elbow connector. Hang
breathing bag onto the hook.
2 Slip breathing hoses with the large cuffs onto the
inspiratory and expiratory ports and connect the small
cuffs to the Y-piece.
3 Connect the sample line to the Y-piece and water
trap.
●
Set the sample rate to 200 mL/min, page 96.
●
Increase tidal volume V
T
to compensate for the
sample gas flow.
Volumeter function:
●
Select "child" scale, page 80.
Self-test and leak test:
4 Firmly connect the Y-piece to the cone.
●
Determine the system compliance and leak-test,
see page 58.
When using CMV mode:
The system compliance will automatically be compen-
sated to ensure that the delivered tidal volume corre-
sponds to the volume setting.
●
Use PCV or CMV mode.
Operation
Ventilating Children
Operating Instructions Julian SW 3.n
59
3
2
1
4
5
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...