Example EtCO
2
:
Upper alarm limit (42 mmHg) and lower alarm limit
(30 mmHg) are shown alongside the measured value
(38 mmHg).
A disabled alarm limit is indicated by two dashes
(– –).
Disabled CO
2
alarms are indicated by a
g
symbol in
the measurement field.
If only the etCO
2
HIGH alarm or the FiCO
2
HIGH
alarm are disabled (in the alarm limit menu), this is
indicated by a
_
symbol in the measurement field.
If APNEA FLOW alarm and MV LOW alarm are dis-
abled, a
g
symbol is displayed in the measurement
field.
If only the MV LOW alarm is disabled, this is indicat-
ed by a
_
symbol.
To set an alarm limit:
●
Place cursor bar on the alarm limit by turning dial
knob, press to confirm.
The alarm limit is highlighted.
●
Set a new value by turning dial knob,
press to confirm.
The new alarm limit is now effective. The cursor bar
returns to the
z
z
z
z
symbol.
For adjustment range of the alarm limits, see table to
the right:
To quit the alarm limits menu:
●
Press the dial knob,
or
●
press the
Q
key.
Monitoring
The Alarm System
Operating Instructions Julian SW 3.n
71
67
38
50
42
30
imits
alarm
info
et
CO
2
Alarm limits
Adjustment range
SpO
2
[%]
_
51 to 100
50 to 99
Pulse
ª
[1/min]
_
31 to 300
20 to 299
EtCO
2
[mmHg]
_
1 to 75
0 to 74
FiCO
2
[mmHg]
_
0 to 10
– –
MV
[L/min]
_
0.1 to 39
0 to 38.9
FiO
2
[Vol.%]
_
19 to 100
18 to 99
Fi Hal.
[Vol.%]
_
0.1 to 7
0 to 6.9
Fi Iso.
[Vol.%]
_
0.1 to 7
0 to 6.9
Fi Enf.
[Vol.%]
_
0.1 to 7
0 to 6.9
Fi Des.
[Vol.%]
_
0.1 to 21.9
0 to 21.8
Fi Sev.
[Vol.%]
_
0.1 to 9.9
0 to 9.8
PAW
[cmH
2
O]
_
5 to 98
4 to 35
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...