Complementary "menu" keys with variable functions
are provided at the bottom edge of the screen, above
each group of "hard" keys. These menu keys are
used to set fresh gas delivery and ventilation
parameters.
1
Left-hand block:
Keys for setting the O
2
concentration and fresh gas
flow.
2
Right-hand block:
Keys for setting parameters for the respective
ventilation mode. The example shows parameters for
controlled, mandatory ventilation (CMV).
These menu keys have different functions, depending
on operating status or ventilation mode.
Current parameter values are displayed in the menu
key field.
In a prominent position at the bottom right-hand side:
The "turn-and-push" dial knob serves as
the main control element of the system and has the
following functions in all setting operations:
3
select = turn
4
confirm = press
–
to confirm the selected carrier gas or a ventilation
mode,
–
to set and confirm parameters for fresh gas and
ventilation modes,
–
to set and confirm monitoring functions.
Next to the dial knob:
The
E
key for switching to standby mode.
5
Press
E
and confirm by pressing dial knob.
Operating Concept
Design of the User Interface
Operating Instructions Julian SW 3.n
21
.
12...
.
N O
2
AIR
MAN
SPONT
CMV
PCV
SpO
2
CMV
alarm
limits
alarm
info
list
curves
config.
SpO
2
EtCO
2
38
98
67
6.0
10
MV
600
V
T
mL
10
10
0
PEEP
cmH
2
O
Freq.
bpm
T
IP
:T
I
%
25
P
max
cmH
2
O
Fresh Gas O
2
+N
2
O
4.00
O
2
%
L/min
50
1
2
T
I
:T
E
Freq
1:2.0
O
2
45
41
2
N O
53
51
Fi
Fet
Iso.
0.8
0.6
Enf.
0.6
0.5
MAC (age 36)
1.3
2
CO
40
0
PAW
0
20
auto set
limits
.
12...
.
N O
2
AIR
MAN
SPONT
CMV
PCV
SpO
2
CMV
alarm
limits
alarm
info
list
curves
config.
SpO
2
EtCO
2
38
98
67
6.0
10
MV
600
V
T
mL
10
10
0
PEEP
cmH
2
O
Freq.
bpm
T
IP
:T
I
%
25
P
max
cmH
2
O
Fresh Gas O
2
+N
2
O
4.00
O
2
%
L/min
50
3
5
4
Freq
T
I
:T
E
1:2.0
O
2
45
41
2
N O
53
51
Fi
Fet
Iso.
0.8
0.6
Enf.
0.6
0.5
MAC (age 36)
1.3
2
CO
40
0
PAW
0
20
auto set
limits
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...