Selecting the Data Screen Page
●
Press
S
key several times until the data page
appears.
Display (example):
Compilation of all numerical values with their units of
measurement. The bar graph on the left of the screen
shows the change in airway pressure PAW over time.
The lines cutting across the bar graph mark upper
and lower alarm limits. To set alarm limits for PAW,
see page 72.
The system compliance and leakage are displayed in
the right-hand field, together with the time of the last
test.
Selecting the Trend Page
Displays a graphical trend of measured values since
measurement was started.
Maximum storable time: 8 hours.
The following display combinations can be selected:
– CO
2
/MV
–
AGas/N
2
O
–
O
2
/compliance
–
SpO
2
/pulse rate (optional)
NOTE: Lung properties can be monitored
continuously with the compliance trend display so
that ventilation can be adjusted accordingly.
Patient compliance is not measured in SPONT
ventilation mode.
●
Press
S
(toggle) key several times until the trend
screen appears.
The scaling of the MV and Compliance trend traces
depends on the »scale« setting in the configuration
menu (pages 78 and 93, Selecting the Scale).
Scale for MV trend:
0 - 3
(child),
0 - 15
(adult)
Scale for compliance trend: 0 - 20
(child),
0 - 100 (adult)
Display (example):
Trend CO
2
and minute volume
Selecting other trend data
●
Press the appropriate menu key:
»AGas/N
2
O«
or
»O
2
/compl.«
or
»SpO
2
/pulse« (optional)
The respective menu key will not show if a measuring
function is not available.
Monitoring
Selecting the Basic Screen Pages
Operating Instructions Julian SW 3.n
67
CMV
PAW
40
20
0
98
67
6.0
0.60
10
27
20
5
15
65
MV
V
T
Freq.
PEAK
PLAT
PEEP
MEAN
Compliance
cmH O
2
cmH O
2
cmH O
2
cmH O
2
mL/cmH O
2
L/min
L
34
AW-Temp.
∞
C
bpm
SpO
2
%
1/min
600 10
10
0
25
1.00
50
alarm
limits
auto set
limits
alarm
info
list
config.
O %
2
L/min
PEEP
cmH
2
O
Freq.
bpm
V
T
mL
T :T
IP
I
%
P
max
cmH
2
O
Fresh Gas O + N O
2
2
1:2.0
15. Dec. 99
sys. comp. 6.40 mL/mbar
leckage 10 mL/min
01-10
8 : 00
12 : 00
T
I
:T
E
N
2
O %
Iso. %
40
33
58
56
0.8
0.6
0.4
0.5
Fi
Fet
O
2
%
Enf. %
1.3
MAC (age: 36)
0
36
CO
2
mmHg
CMV
98
67
600 10
10
0
25
1.00
50
PAW
40
20
0
11:00
12:00
13:00
60
30
0
15
10
5
0
To select zoom area:
Turn !
To zoom in: Press
!
CO
2
MV
SpO
2
alarm
limits
agent
N O
2
CO
2
MV
O
2
compl.
SpO
2
pulse
full
trend
O %
2
L/min
PEEP
cmH O
2
Freq.
bpm
V
T
mL
T :T
IP
I
%
P
max
cmH O
2
Fresh Gas O + N O
2 2
et
CO
2
38
6.0
10
MV
Freq
T
I
:T
E
1:2.0
O
2
45
41
2
N O
53
51
Fi
Fet
Iso.
0.8
0.6
Enf.
0.6
0.5
MAC (age: 36)
1.3
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...