![Dräger Julian Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787050.webp)
CMV Ventilation Mode
CMV = Continuous Mandatory Ventilation
Volume-controlled ventilation with fixed mandatory
minute volume (MV), set with the tidal volume (V
T
),
and the respirtory rate (Freq.).
Presetting the ventilation parameters for CMV
1 Press »CMV« key; its LED will start flashing.
●
The six ventilation parameters for CMV are displayed
on the screen with their default settings.
2 Press the menu key for the desired ventilation
parameter.
3 Set and confirm the ventilation parameter with the dial
knob.
Display (example):
Adjustment ranges and factory default setting
For tidal volumes V
T
of less than 200 mL:
●
Use a pediatric patient circuit, see page 59
"Ventilating Children".
____
*
User-specific defaults may be selected, see page 100.
Ventilation parameters
Adjustment
range
Factory default
setting *
Pressure limit
Pmax [cmH
2
O]
10 to 70
25
Tidal volume V
T
[mL]
50 to 1400
600
Frequency Freq.
[1/min]
6 to 60
12
Insp/Exp ratio
T
I
: T
E
2.0:1 to 1:2.0 in
increments of 0.1,
1:2.0 to 1:4.0, in
increments of 0.5
1 : 2.0
Insp. pause time :
Insp. time
T
IP
: T
I
[%]
0 to 50
10
PEEP [cmH
2
O]
0 to 20
0
Operation
Selecting Ventilation Modes
Operating Instructions Julian SW 3.n
50
Julian
2
3
1
delete
trend
test
config.
Julian
Standby
To Start CMV: Confirm !
leak
03.00
01. May 00
1:2.0
T
I
:T
E
4.00
50
12
10
0
25
O %
2
L/min
PEEP
cmH O
2
Freq.
bpm
T :T
IP
I
%
P
max
cmH O
2
Fresh Gas O +N O
2
2
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...