![Dräger Julian Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787042.webp)
If all points on the checklist are OK:
●
Confirm = press dial knob.
The self test is started.
Duration: 3 to 4 minutes.
The self test is performed automatically.
1 Julian essentially carries out the following automatic
tests and actions, which are coded:
The following list gives an overview of the main steps
of the self test:
Electronics
– gas analyzer test and calibration
(incl. Monitoring) – O2-sensor test and calibration
– PAW-sensor test and calibration
– flow sensor check
– backup battery check
– restoring of default settings for
alarm limits, monitoring parameters,
ventilator settings, and fresh gas
flow control settings
Fresh Gas
– gas supplies check(O2, Air, N2O)
Flow Control
– gas inlet valves check
– pressure sensors calibration
– flow control valve calibration
– pressure release valve test
– check for closed Safety O
2
control
valve
– fresh gas shortage detection check
Ventilator
– Testing and calibration of
PEEP/Pmax valves
– Testing of safety valve
– Testing of MAN/AUTO switch-over
valve
– Testing of breathing phase valve
– Testing of internal pressure sensor
– Determination of compliance and
leakage
NOTE: If, during the self test, any malfunction or missing
gases are detected, a respective advisory will be
displayed on the screen.
In some cases, the user has a choice of accepting the
malfunction or of repeating that part of the self test
(example: AIR supply not connected).
In other cases, the user has to remedy the problem
before the self test can continue (example: O
2
supply not
connected).
Preparation
Checks of Readiness for Operation
Operating Instructions Julian SW 3.n
42
Self-test
electronics
freshgas
mixer
ventilator
✓
✓
cancel
test
24
1
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...