![Dräger Julian Скачать руководство пользователя страница 73](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787073.webp)
CO
2
Alarms On/Off
The alarm limits for EtCO
2
, insp. CO
2
and CO
2
apnea
monitoring can be switched off with the
menu key »CO2-al. on/off« in MAN/SPONT mode.
In this case, the CO
2
g
symbol appears alongside
the measured endexpiratory CO
2
concentration.
The CO
2
alarm limits are reactivated by pressing the
menu key again. The disabled CO
2
alarm limits are
reactivated automatically when changing to a different
ventilation mode.
The CO
2
alarm limits can also be enabled and
disabled under "Alarm limits" in the configuration screen
in all ventilation modes, see page 81.
Disabled CO
2
alarms are indicated by a
g
symbol in the
measurement field.
If only the etCO
2
HIGH alarm or the FiCO
2
HIGH alarm
are disabled (in the alarm limit menu), this is indicated by
a
_
symbol in the measurement field.
The etCO2
_
and FiCO2
>
alarms can be configured
»ON« or »OFF« in standby mode for the transition from
automatic ventilation. When the alarm limits are set to
»ON«, the value is transfered from automatic ventilation
mode.
Alarm Info
This function is used to list all active alarms in order of
priority.
●
Press and hold »alarm info« menu key.
Display (example):
Active WARNING level alarm:
FIO2
<
<
<
<
!!!
Active CAUTION-level alarm:
ETCO2
<
<
<
<
!!
Active ADVISORY-level alarm:
MIXED AGENT !
The display only remains visible as long as the menu key
is pressed.
Up to 8 alarms can be displayed simultaneously.
Monitoring
The Alarm System
Operating Instructions Julian SW 3.n
73
CMV
600
10
10
0
25
1.00
50
ETCO2
MIXED AGENT
!!
!
FIO2
!!!
alarm
limits
auto set
limits
alarm
info
list
curves
config.
O %
2
L/min
PEEP
cmH O
2
Freq.
bpm
V
T
mL
T :T
IP
I
%
P
max
cmH O
2
Fresh Gas O + N O
2
2
Pleth.
1:2.0
T
I
:T
E
CO
2
40
0
PAW
0
20
MAN/SPONT
98
67
4.00
30
To start Volumeter, confirm !
SpO
2
alarm
limits
alarm
info
list
curves
config.
Fresh Gas O + N O
2
2
O %
2
L/min
V
T
0.35
0
Volumeter
--s
--
0
5
10
0.5
1
et
CO
2
38
6.0
10
MV
Freq
CO al.
2
on/off
Pleth.
g
CO
2
O
2
45
41
2
N O
53
51
Fi
Fet
Iso.
0.8
0.6
Enf.
0.6
0.5
MAC (age 36)
1.3
g
MV
CO
2
40
0
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...