Audible alarms can be silenced for two minutes:
NOTE: A new alarm condition during this period will
still be signaled with an audible alarm.
1 Press
G
key and its yellow LED will light up.
To reactivate audible alarm:
1 Press
G
key and its yellow LED will go out.
A list of alarm messages can be found on page 122,
"Troubleshooting".
CO
2
alarms and SpO
2
alarms are disabled after
switching from standby to a ventilation mode. They
are automatically activated after the first correspon-
ding signal has been detected.
Displaying and Setting Alarm Limits
Alarms can be displayed and set from all three basic
screens (standard, data and trend screen) during
operation.
The default alarm limits for the ventilation mode are
activated automatically when changing from standby
to a ventilation mode.
These default alarm limits can be used unchanged or
they can be set individually for a patient (see page 98,
"Setting Default Alarm Limits").
Individual settings are deleted on changing from a
ventilation mode to standby when default alarm limits
are reactivated.
To call up alarm limits:
●
Press the menu key »alarm limits«.
Display (example):
Measured values and their alarm limits are shown on
the right-hand side.
Waveforms with alarm limits represented by dashed
lines are shown on the left.
WARNING !
The alarm silence button is intended to provide a
way of muting audible alarms while corrective
action is taken. Failure to identify and correct
alarm situations may result in patient injury.
Monitoring
The Alarm System
Operating Instructions Julian SW 3.n
70
Pleth
CMV
600 10
10
0
25
1.00
50
98
67
38
6.0
29
0.8
3.0
--
20
1.0
0.0
40
8
--
To adjust alarm limit, turn and confirm !
FiO
2
FiHal.
PAW
SpO
2
alarm
limits
auto set
limits
alarm
info
list
config.
O %
2
L/min
PEEP
Freq.
V
T
mL
T :T
IP
I
%
P
max
Fresh Gas O + N O
2 2
curves
et
CO
2
MV
cmH
2
O
bpm
cmH
2
O
--
95
120
50
42
30
1:2.0
T
I
:T
E
CO
2
40
0
PAW
0
20
Pleth
CMV
600 10
10
0
25
1.00
50
98
67
38
6.0
29
0.8
--
95
3.0
--
20
1.0
0.0
40
8
120
50
--
To select alarm limit, turn and confirm !
42
30
FiO
2
FiHal.
PAW
SpO
2
alarm
limits
auto se
limits
alarm
info
list
config.
O %
2
L/min
PEEP
Freq.
V
T
mL
T :T
IP
I
%
P
max
Fresh Gas O + N O
2 2
curves
et
CO
2
MV
cmH
2
O
bpm
cmH
2
O
1:2.0
T
I
:T
E
CO
2
40
0
PAW
0
20
Julian
1
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...