Manual Calibration
Julian calibrates its sensors automatically during
operation. The message CAL appears instead of the
current measured value during the calibration process.
Calibration of the O
2
sensor and zero calibration of
the other gas sensors can be repeated manually, for
instance if automatic calibration was unsuccessful.
In standby/configuration mode:
●
Turn dial knob to select »calibration« column,
press to confirm.
●
Select and confirm appropriate line with dial knob,
for example:
»O
2
-sensor
21 vol.%«
The calibration process is started when the selection
is confirmed. The clock symbol
U
appears at the end
of the respective line, indicating that calibration is in
progress.
When calibration is complete, a tick mark (
✓
) appears
instead of the clock symbol. A question mark (?)
indicates that calibration must be repeated because
the previous calibration was not completed success-
fully.
To further improve measuring accuracy of high O
2
concentrations, an additional 100 vol% O
2
calibration
and a linearity test of the O
2
sensor can be performed
under the »more« menu option.
Other settings can already be made while calibration
is in progress.
100 Vol.% O
2
Calibration
●
Use a separate O
2
source, e.g. O
2
flowmeter.
O
2
from the Julian breathing system is not suitable, as
it will still contains traces of anesthetic gas.
●
Unscrew sample line from the Y-piece and place it in
the continuous flow of the O
2
source.
●
Turn dial knob to select the choice
»O
2
-sensor
100 vol.%«
and press to confirm.
●
Allow continuous flow of O
2
into the sample line.
A check mark (
✓
) appears after the field when
calibration is complete.
●
Screw the sample line back onto the Y-piece.
Configuring in Standby Mode
Manual Calibration
Operating Instructions Julian SW 3.n
102
Standby / Configuration
calibration
defaults
O -sensor
2
21 vol.%
more
✓
alarm sound
scale
parameter
record
interface
alarm limits
curves
basic configurations
CMV settings
PCV settings
Fresh Gas Flow Control
0 1 2 3 4 6 7 8 9
5
1 2 3 5 6 7 8 9
4
gas
sensor
pls. tone
adult
child
✓
MAC-config.
MAC display
on
off
MAC correction
on
off
age 36
✓
✓
Standby / Configuration
O
2
-sensor
21 Vol.%
gas sensor
more
O
2
-sensor
100 Vol.%
linearity
test O
2
calibration
Confirm !
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...