![Dräger Julian Скачать руководство пользователя страница 104](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787104.webp)
Displaying MAC
To activate the MAC display:
●
Select column »MAC-config.« by turning the rotary
knob and press to confirm.
●
Select line »MAC display« by turning the rotary knob
and press to confirm.
●
Select line »on« by turning the rotary knob and press
to confirm.
To select the age correction for the MAC display:
●
Select line »MAC correction« by turning the rotary
knob and press to confirm.
●
Select line »on« by turning rotary knob and press to
confirm.To select the patient's age:
●
Select line »age« by turning the rotary knob and press
to confirm.
●
Set age by turning the rotary knob and press to confirm.
If the MAC display is configured to »off«,
the »MAC correction« and »age« lines are
automatically dimmed.
If MAC correction is configured to »off«, the »age«
line is automatically dimmed.
MAC-Definition:
1 MAC is equal to the anesthetic gas concentration at
which 50% of all patients no longer respond to a
stimulation of the nerves.
The influence of other medication(opiates or intrave-
nous hypnotics) is not taken into consideration when
calculating MAC values.
Monitoring
MAC configuration
Operating Instructions Julian SW 3.n
104
600 10
10
0
25
1.00
50
O %
2
L/min
PEEP
cmH O
2
Freq.
bpm
V
T
mL
T :T
IP
I
%
P
max
cmH O
2
Fresh Gas O + N O
2
2
PAW
40
20
0
Confirm to select Isoflurane.
O
2
-Sensor
21 Vol.%
gas sensor
CMV
calibration
pulse tone
alarm sound
scale
parameters
record
activate monitor defaults
0 1 2 3 4
5
6 7 8 9
1 2 3
4
5 6 7 8 9
adult.
child
alarms
settings
✓
✓
T
I
:T
E
1:2.0
default
CO
2
on
off
SpO
2
on
off
HLM on
off
FiCO
2
5
MAC display
on
off
MAC correction
on
off
age 36
MAC-config.
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...