In the table below, messages are listed in alphabetical
order.
The table should help you to identify the cause of an
alarm and to ensure rapid remedy of the problem.
Troubleshooting
Julian displays alarm messages at 3 levels of priority:
(For definitions of alarm priority levels, see page 69)
!!! Warning = Message with top priority
!!
Caution
= Message with medium priority
!
Advisory = Message with low priority
Troubleshooting
Operating Instructions Julian SW 3.n
122
Message
Cause
Remedy
AGT ERR
!
* Anesthetic gas measurement faulty.
Call DrägerService.
2
nd
AGENT
!
2
nd
Anesthetic agent recognized
Flush system if necessary.
Check fresh gas setting.
Wait for transitional phase.
AIR SUPPLY
!!!
Compressed air supply has failed.
Central supply connector not plugged
in.
Check connection to central supply.
Compressed air hose kinked.
Air compressor failure.
Check compressor.
APNEA CO
2
!!!
Breathing/ventilation has stopped.
No breath detected for 15 seconds.
WARNING Patient must immediately
be ventilated by hand!
Check ventilator setting. Ensure it is not
disconnected.
No breath detected for 60 seconds in
"MAN/SPONT alarm limits" alarm
mode.
WARNING Patient must immediately
be ventilated by hand!
Check patient's spontaneous breathing
ability.
APNEA PRESSURE
!!!
Ventilation has stopped.
No change of pressure detected for
15 seconds.
WARNING Patient must immediately
be ventilated by hand!
Inadequate supply of fresh gas.
Check fresh gas setting.
Leak in patient circuit.
Check that hose system is not
disconnected.
Patient connection open.**
Connect patient up correctly.
APNEA VOL
!!!
Ventilation has stopped.
No expiratory tidal volume for
15 seconds.
WARNING Patient must immediately
be ventilated by hand!
Check ventilator setting.
Inadequate supply of fresh gas.
Check fresh gas setting.
Circuit kinked.
Check hose system.
*
Advisories related to hardware errors (e.g. sensor failures) are announced without audible signal.
** In automatic ventilation mode and with PEEP = 0, due to its very
low internal breathing resistances, the workstation can display a
measured value for the minute volume. This display is irrelevant in
conjunction with the message "APNEA PRESSURE!!!".
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...