![Dräger Julian Скачать руководство пользователя страница 115](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787115.webp)
The parts should preferably be cleaned and disinfected in a parts washing/disinfection unit, or by immersion.
Julian and its components must not be treated with formaldehyde vapors or ethylene oxide!
115
Sterilization
Washing machine
2)
93
°
C 10 minutes
Wiping
3)
Immersion
3)
Steam, 134
°
C
(273 ºF), 10 min.
no
outside
no
no
no
outside
no
no
no
yes
no
no
yes
yes
no
yes
yes
no
possible
yes, max. 25 times
yes
no
possible
yes
yes
no
possible
yes
yes
no
possible
yes
yes
no
possible
yes
yes
no
possible
yes
yes
no
possible
yes
yes
no
possible
yes
no
outside
4)
no
yes, max. 10 times
no
outside
possible
5)
no
yes
no
possible
no
no
yes
no
no
yes
no
possible
yes
yes
no
possible
no
yes
no
possible
no
yes
no
possible
no
yes
no
possible
yes
2) Only use cleaning agents, no alkaline or chlorine-releasing substances. Risk of corrosion!
3) Surface disinfectants based on aldehydes, alcohols and quaternary ammonium compounds, such
as PaMo dur rapid or PaMo sept (Dräger Parts&More GmbH) can be used to ensure compatibility.
4) Do not use any agents containing alcohol.
5) Disinfect the flow sensor in 70% ethanol solution for approximately1 hour and leave to dry in air for
at least 30 minutes.
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...