Troubleshooting
Operating Instructions Julian SW 3.n
125
Message
Cause
Remedy
N
2
O SUPPLY ?
!!!
N
2
O supply has failed.
Open the N
2
O backup cylinder.
Supply pipeline connector is not
plugged in or N
2
O hose is kinked.
Check connection to supply pipeline.
Restore N
2
O supply.
N
2
O cylinder empty.
Connect a full N
2
O cylinder.
O
2
SUPPLY ?
!!!
Oxygen supply has failed.
Open the O
2
backup cylinder.
The supply pipeline connector is not
plugged in or O
2
hose is kinked.
O
2
cylinder empty.
Check connection to supply pipeline.
Restore O
2
supply.
O
2
cylinder empty.
Connect a full O
2
cylinder.
PAW HIGH
!!!
Airway pressure exceeds upper alarm
limit.
Ventilation hose kinked.
Check hose system.
Stenosis.
Pressure limit set too high.
Correct pressure limit.
PAW NEGATIVE
!!!
Inadequate supply of fresh gas.
Set adequate fresh gas flow on the
anesthesia workstation
Mean pressure p
mean
less than
–2 cmH
2
O.
Airway pressure PAW less than
–7 cmH
2
O least 3 seconds.
P MAX?
!
PCV pressure limit not reached.
Seal the leak.
Correct pressure limit, inspiratory
flow or inspiration time if necessary.
POWER FAIL
!
Electrical supply failure.
Check battery capacity.
Prepare manual ventilation.
PRESS ERR
!
* Pressure sensor failed.
Call DrägerService.
PRESSURE LIM
!!
The ventilator is operating in pressure-
limited mode.
Lung compliance has changed.
Tube kinked.
Bacteria filter on inspiratory or expiratory
side is clogged .
Check tube/bacteria filter.
If necessay, increase P
max
or
reduce V
T
.
SAFETY O
2
ON
!!
Safety O
2
control valve is open during
normal operation.
Close Safety O
2
immediately.
SPO
2
ERR
!
*
SpO
2
measurement defective.
Call DrägerService.
SPO
2
PULS LO
!!!
Pulse rate has dropped below set alarm
limit.
Check patient's condition!
SPO
2
PULS HI
!!
Pulse rate exceeds upper alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
* Advisories related to hardware errors (e.g. sensor failures) are announced without audible signal.
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...