Connecting Supplies
Gas Connections
Connecting to pipeline gas supply
1 Screw the gas supply pipeline delivery hoses for O
2
,
AIR, N
2
O, and VAC into the respective inlet ports on
the back of the gas supply block. (The two ports at
the front of the gas supply block are reserved for the
backup gas cylinders.)
Connect the other end of the high pressure gas deli-
very hoses to the wall or ceiling gas outlets.
The table below shows color codes of gas
connections:
●
Ensure that the gas pressures of the gas supply
pipeline are between 270 and 550 kPa
(39 psi and 80 psi):
2 All pipeline pressure displays must be in the specified
range.
The pressure displays light up when the Julian
workstation is connected to line power.
USA
Canada
O
2
green
white
AIR
yellow
black/white checkered
N
2
O
blue
blue
VAC
magenta
magenta
WARNING !
Each time you connect a machine to a wall or
ceiling gas outlet carefully check hoses to
ensure that both ends of the hose are indexed
for the same gas. Pipeline delivery hoses used
between wall outlets and anesthesia
machines have caused accidents when,
during assembly, an oxygen fitting was placed
on one end of the hose and a nitrous oxide
fitting on the other end.
WARNING !
Always use medical grade gases that are dry
and free from dust and oil. Contaminated gas
may cause system malfunction.
Preparation
Connecting Supplies
Gas Connections
Operating Instructions Julian SW 3.n
33
N
2
O AIR O
2
VAC
1
Julian
D
2
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...