Interface
Choice between configuration of COM2 or COM3.
Baud rate
Transmission rate (variable, see Instructions for Use
of the equipment to be connected).
Parity
This display is fixed for MEDIBUS and is shown for
information purposes only.
Data bits
This display is fixed for MEDIBUS and is shown for
information purposes only.
Stop bits
This display is fixed for MEDIBUS and is shown for
information purposes only.
Setting Default Alarm Limits
●
Turn the rotary knob to select the line »alarm limits«
and press to confirm.
The »alarm limits« menu appears in the left-hand
field.
Display (example):
●
Select required parameter by turning the rotary knob
and press to confirm.
The "alarm limits" menu appears.
●
In the menu, select the required alarm limit by turning
the rotary knob and press to confirm. The default
alarm limit is highlighted against a white background.
●
Set and confirm the value with the rotary knob. The
value is then displayed in normal mode.
The cursor bar jumps to the next default alarm limit.
●
Set the next default alarm limit as described above.
The following alarm limits can be configured:
The etCO
2
_
, FiCO
2
>
, FiO
2
>
and MV
_
alarms
can be configured »ON« or »OFF« in standby mode for
the transition from MAN/SPONT. When the alarm limits
are set to »ON«, the value is transferred from automatic
ventilation mode.
Configuring in Standby Mode
Setting Default Values
Operating Instructions Julian SW 3.n
96
Standby / Configuration
MAC-config.
calibration
0 1 2 3 4
5
6 7 8 9
1 2 3
4
5 6 7 8 9
adult
child
default settings
pulse tone
alarm sound
scale
parameters
record
interfaces
alarm limits
curves
basic configurations
IPPV default settings
PCV default settings
fresh gas delivery
SpO
2
120
50
90
100
SpO
2
/pulse
CO
2
/O
2
Flow/pressure
Anaesthetic agent
MAN/SPONT alarms
SpO
2
/pulse
SpO
2
CO
2
/O
2
etCO
2
FiCO
2
FiO
2
Flow/pressure MV
PAW
Anesth. agent
Fi Iso. Fi Hal.
Fi Sev.
Fi Enf. Fi Des.
MAN/SPONT
alarms
etCO
2
FiCO
2
FiO
2
MV
Pulse rate
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...