![Dräger Julian Скачать руководство пользователя страница 123](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787123.webp)
Troubleshooting
Operating Instructions Julian SW 3.n
123
Message
Cause
Remedy
AW TEMP HIGH
!!!
Breathing gas temperature above
40
°
C.
If using a breathing gas humidifier:
disconnect humidifier.
If using heated patient circuit:
remove plug of circuit heater.
BATTERY LOW
!!!
Battery capacity of the uninterruptible
power supply is almost exhausted.
Prepare manual ventilation with
100 % O
2
.
CO
2
/AGT ERR
!
* CO
2
/ anesthetic gas measurement
faulty, and consequently:
aspiratory O
2
measurement faulty.
Call DrägerService.
CO
2
ALRM OFF
!
*
APNEA CO2, INSP CO2 HI or
ET CO2 HIGH alarm disabled.
CO
2
ERR
!
* CO
2
gas measurement faulty.
Call DrägerService.
CO
2
LINE BLK
!
Sample line blocked.
Check sample line and water trap:
replace if necessary.
Check that hose to anesthetic gas
scavenging system is not kinked.
COOLING ?
!
* Internal system temperature is too high
or fan for mixer electronics is defective.
Clean filter on the back panel.
Call DrägerService.
ET CO
2
HIGH
!!
Upper alarm limit for end-expiratory
CO
2
concentration has been exceeded
for at least two breaths.
Check ventilation.
ET CO
2
LOW
!!
End-expiratory CO
2
concentration has
fallen below lower alarm limit for at least
two breaths.
Check ventilation.
EXP MAC
>
!!
Upper alarm limit for expiratory MAC
value has been exceeded.
Check Vapor setting.
Check fresh gas setting.
% HAL HIGH
!!!
% ISO HIGH
!!!
% ENF HIGH
!!!
% DES HIGH
!!!
% SEV HIGH
!!!
Inspiratory gas concentration of the
respective anesthetic agent exceeds
upper alarm limit.
Upper alarm limit has been exceeded
for at least two breaths.
Check setting of anesthetic
vaporizer.
% HAL LOW
!!
% ISO LOW
!!
% ENF LOW
!!
% DES LOW
!!
% SEV LOW
!!
Inspiratory gas concentration of the
respective anesthetic agent is below
lower alarm limit.
Value has remained below the lower
alarm limit for at least two breaths.
Check setting of anesthetic
vaporizer.
% O
2
LOW
!!!
Inspiratory O
2
concentration is below
lower alarm limit.
Check O
2
supply.
Check O
2
concentration in fresh gas
flow.
* Advisories related to hardware errors (e.g. sensor failures) are announced without audible signal.
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...