Anesthetic gas measurement
All data in % by volume at
(infrared spectrometry)
1013 mbar current atmospheric pressure
Sample rate (selectable)
60 mL/min or 200 mL/min
Delay time for sampling
Less than 1 s (for a sample rate of 200 mL/min)
Response time t 10...90
at 200 mL/min
300 ms
at 60 mL/min
1 s
Accuracy
The measurement accuracy required by DIN ISO 11196 is
attained at the latest 4 minutes after switching on the apparatus.
The threshold for automatic detection is 0.15% by volume.
To check the complete accuracy of Julian, a temperature
stabilisation time of 1 hour at 21
°
C should be observed.
The accuracy of the measuring equipment used should be at least
5 times and if possible 10 times higher than that of Julian.
Display range for N
2
O
0 to 100 vol%
Resolution
0.1 vol%
Accuracy
Better than
±
2 vol% absolute or
±
3 % of measured value
Display range for halothane
0 to 8.5 vol%
Resolution
0.1 vol%
Accuracy for
0 to 4 vol%
0.15 vol% or 5 % of measured value
>4 to 8.5 vol%
±
10 % of measured value
Display range for isoflurane
0 to 8.5 vol%
Resolution
0.1 vol%
Accuracy for
0 to 5 vol%
0.15 vol% or 5 % of measured value
>5 to 8.5 vol%
±
10 % of measured value
Display range for enflurane
0 to 9.9 vol%
Resolution
0.1 vol%
Accuracy for
0 to 54 vol%
0.15 vol% or 5 % of measured value
>4 to 9.9 vol%
±
10 % of measured value
Display range for sevoflurane
0 to 9.9 vol%
Resolution
0.1 vol%
Accuracy for
0 to 6 vol%
0.15 vol% or 5 % of measured value
>6 to 9.9 vol%
±
10 % of measured value
Display range for desflurane
0 to 22 vol%
Resolution
0.1 vol%
Accuracy for
0 to 15 vol%
0.15 vol% or 5 % of measured value
>8 to 22 vol%
±
10 % of measured value
Automatic anesthetic gas identification
Volatile anesthetic agents are automatically identified.
Manual selection is not necessary and not possible. In the case
of a mixture of two volatile anesthetic agents, the concentration of
the secondary anesthetic agent is displayed from a MAC value of
0.1 MAC. A secondary anesthetic agent becomes the main
anesthetic agent if its MAC value exceeds the MAC value of the
main anesthetic agent by 0.2 MAC. If a mixture of more than
2 volatile anesthetic agents is present, the display is blank.
Cross-sensitivity
None with respect to alcohol, acetone, methane, NO and CO.
Technical Data
Measuring Functions
141
Operating Instructions Julian SW 3.n
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...