![Dräger Julian Скачать руководство пользователя страница 81](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787081.webp)
Configuring Alarms
●
Turn dial knob to select the column »alarms«, press
to confirm.
Default
This function is used to activate default alarm limits
set in the standby configuration which are effective
whenever the system is switched to standby.
Setting default alarm limits:
●
Turn dial knob to select »defaults« menu, press to
confirm.
Please refer to page 96 for adjusting default limits.
CO
2
alarms on/off
To enable/disable all CO
2
alarms, including the apnea
alarm.
●
Turn dial knob to select »CO
2
on« or »CO
2
off«, press
to confirm.
Disabled CO
2
alarms are indicated by a
g
symbol in
the measurement field.
If only the etCO
2
HIGH alarm or the FiCO
2
HIGH
alarm are disabled (in the alarm limit menu), this is
indicated by a
_
symbol in the measurement field.
SpO
2
alarms on/off
To enable/disable all SpO
2
alarms (optional).
●
Turn dial knob to select »SpO
2
on« or »SpO
2
off«,
press to confirm.
Disabled SpO
2
alarms (except SPO2 SEN Disc and
SPO2 ERR) are indicated by the
_
symbol in the
measurement field.
Deactivated alarms are indicated by the symbol
_
in the measurement field with a corresponding
message.
If all the alarms for this measured value are
deactivated, this is indicated by the symbol
G
and
a corresponding message.
Monitoring
Configuring During Operation
Operating Instructions Julian SW 3.n
81
CMV
600 10
10
0
25
1.00
50
O %
2
L/min
PEEP
cmH
2
O
Freq.
bpm
V
T
mL
T :T
IP
I
%
P
max
cmH
2
O
Fresh gas O + N O
2
2
PAW
40
20
0
O
2
-Sensor
21 Vol.%
gas sensor
calibration
pulse tone
alarm sound
scale
parameters
record
activate monitor defaults
0 1 2 3 4
5
6 7 8 9
1 2 3
4
5 6 7 8 9
adult.
child
1:2.0
✓
✓
T
I
:T
E
settings
alarms
default
CO
2
on
off
SpO
2
on
off
HLM on
off
FiCO
2
5
MAC-config.
MAC display
on
off
MAC correction
on
off
age 36
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...