Certain alarms are automatically switched off in
MAN/SPONT ventilation mode in order to avoid nuisance
alarms. Disabled alarms are identified by a gray back-
ground in the table below.
– – : The alarm limit is deactivated as factory default.
*
The EtCO
2
_
, FiCO
2
>
, MV
_
and FiO
2
>
alarms can be configured »ON« or »OFF« in
standby mode for the transition from
MAN/SPONT. When the alarm limits are set to
»ON«, the value is transferred from automatic
ventilation mode.
To set the values of the alarm limits during operation
see page 72.
To configure the default alarm limits, see page 96.
Alarm
Limits
In MAN/SPONT
mode
Factory default
setting
SpO
2
_
ON
ON
– –
92
Pulse
ª
_
ON
ON
120
50
EtCO
2
_
*
*
50 mmHg
– –
FiCO
2
>
*
– –
MV
_
*
*
– –
– –
FiO
2
_
*
ON
– –
20
Fi Hal.
_
ON
OFF
1.5
– –
Fi Iso.
_
ON
OFF
2.3
– –
Fi Enf.
_
ON
OFF
3.4
– –
Fi Des.
_
ON
OFF
12.0
– –
Fi Sev.
_
ON
OFF
3.4
– –
PAW
_
ON
OFF
40
– –
Apnea pressure
OFF
Apnea flow
OFF
Apnea CO
2
ON
active after
60 sec.
Operation
Selecting Ventilation Modes
Operating Instructions Julian SW 3.n
49
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...