![Dräger Julian Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787029.webp)
Assembly of Components
For assembly and installation of the breathing system,
bellows, and absorber system, please refer to
page 117, "Assembly".
Installing the Breathing Circuit
1 Connect the limbs of the breathing hoses to the Y-
piece and to the ports of the breathing system.
2 Connect the 2.3 L breathing bag to the breathing
hose with a connecting sleeve. Plug the breathing
hose onto the angled connector and hang the
breathing bag on the hook.
3 Fit water trap into its holder.
4 Screw the sample line onto the Y-piece and water
trap.
WARNING !
Only use original gas sample line. Other lines
may change the technical data of the device.
WARNING !
Do not use antistatic or conductive patient
breathing hoses!
They present a risk of burn injuries when used
in combination with HF electrosurgery.
WARNING !
Always install parts that have been cleaned
and disinfected!
Preparation
Assembly of Components
Operating Instructions Julian SW 3.n
29
4
3
4
1
1
2
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...