![Dräger Julian Скачать руководство пользователя страница 109](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787109.webp)
Opening the breathing system
– Weekly
1 Loosen the five sealing screws a quarter-turn with the
key supplied.
●
Remove cover.
2 Remove the two valve discs.
●
Lift off the metal valve plate.
●
Keep valve discs in a cassette to protect them against
damage and prepare them for cleaning and disin-
fection in a parts washing/disinfection unit.
●
Prepare housing parts for disinfection and
cleaning in a parts washing/disinfection unit.
Anesthetic Gas Scavenging System (AGS)
3 Remove gas-disposal-assembly tubing.
4 Remove the transfer hose.
5 Remove the anesthetic gas scavenging system.
●
Prepare the individual parts for cleaning and disin-
fection in a parts washing/disinfection unit.
Care
Disassembly
Operating Instructions Julian SW 3.n
109
1
1
1
1
1
2
2
5
4
4
3
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...