PCV Ventilation Mode
PCV = Pressure Controlled Ventilation
In pressure controlled ventilation mode, the applied
tidal volume depends on the ventilation parameters
Pmax and PEEP as well as on lung compliance.
Changes in ventilation parameters or lung compliance
influence tidal volume. Minute volume (MV or V
t
) must
therefore be monitored continuously.
Presetting ventilation parameters for PCV
1 Press »PCV«. Its LED will start flashing.
●
The five ventilation parameters for PCV appear on the
screen with their default settings.
2 Press menu key of the respective ventilation
parameter.
3 Set and confirm ventilation parameter with the dial
knob.
Display (example):
Adjustment ranges and factory default settings
When ventilating infants and small children weighing
less than 20 kg:
●
Use pediatric patient circuit; see page 59.
"Ventilating Children".
–––––––––––
*
User-specific defaults may be selected, see page 102
.
** See technical data, page 140.
Ventilation parameter
Adjustment
range
Factory default
setting *
Pressure limit
Pmax [cmH
2
O]
(PEEP+1)
to 70
15
Breathing rate
(Freq.) [bpm]
6 to 60
12
Insp./Exp. ratio
T
I
: T
E
2.0:1 to 1:2.0 in
increments of 0.1,
1:2.0 to 1:4.0, in
increments of 0.5
1 : 2.0
Insp. flow [L/min]
5 to 50 (75**)
30
PEEP [cmH
2
O]
0 to 20
0
Operation
Selecting Ventilation Modes
Operating Instructions Julian SW 3.n
53
Julian
03.00
01. May 00
Standby
delete
trend
leak
test
config.
To Start PCV : Confirm !
1:2.0
T
I
:T
E
4.00
50
12
30
0
15
O
2
%
L/min
PEEP
cmH O
2
Freq.
bpm
Insp.Flow
L/min
P
max
cmH O
2
Fresh gas O +N O
2
2
1
Julian
2
1
3
Содержание Julian
Страница 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Страница 3: ...XXXXXXX ...
Страница 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Страница 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Страница 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Страница 121: ...Contents Troubleshooting 122 Troubleshooting Contents Operating Instructions Julian SW 3 n 121 Troubleshooting ...
Страница 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Страница 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...