Aespire View
3-14
M1132382
•
The active low flow system with a 25 mm barb connector is for
use with low vacuum disposal systems. It requires an external
venturi/ejector system with a 40 to 50 l/min extract flow. A flow
indicator on the system indicates when the unit is in operation.
•
The active low flow system with 30-mm ISO taper is for use with
low vacuum disposal systems. It requires an external venturi/
ejector system with a 40 to 50 l/min extract flow. A flow indicator
on the system indicates when the unit is in operation.
See “
” for additional scavenging connection
information.
Connecting active
AGSS with a flow
indicator
To use the optional active AGSS on a system that has a flow
indicator, connect it as follows:
1. Connect the proper hose to the AGSS outlet connector on the
bottom of the AGSS, beneath the breathing system. Attach the
other end to the hospital disposal system.
2. With the AGSS operating, verify that the flow indicator ball on the
flow indicator rises to the green zone, indicating adequate flow.
Note
The ball in the upper red zone indicates excessively high extraction
flow. The ball in the lower red zone indicates extraction flow rate is
too low or the filter is blocked.
3. Complete the tests in the
section of this
manual.
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...