Aespire View
8-8
M1132382
3. Lift off the cover to remove it.
4. Remove and properly discard the foam filters, the absorbent, and
any water in the reservoir.
WARNING
Be careful when draining condensate from the absorber.
The liquid is caustic and may burn skin.
5. To clean and disinfect the canister, see
“Absorber canister”
in the
“Cleaning and Sterilization”
manual.
6. Assemble canister.
•
Place a new filter in the bottom of the canister.
•
Pour absorbent into the canister.
•
Place a new filter over the absorbent.
•
Align the cover slots with the canister locking tabs and press
the cover down into place.
•
Turn the cover locking ring clockwise to lock the cover in
place.
•
Ensure cover is properly sealed to prevent leaks and spillage.
•
Wipe off any absorbent dust.
Note
Alignment of the arrows helps to indicate correct assembly.
AC
.2
0p
003
AC.
2
0p
00
0
4