Aespire View
5-8
M1132382
Vaporizer back pressure test
WARNING
Anesthetic agent comes out of the common gas outlet
during this test. Use a safe, approved procedure to
remove and collect the agent.
CAUTION
To prevent damage to the vaporizer, turn the flow controls
fully clockwise (minimum flow or off) before turning on the
system.
1. Set the System switch to On. Alarms may occur.
2. Set the O2 flow to 6 l/min.
3. Make sure that the O2 flow stays constant and the float moves
freely.
4. Adjust the vaporizer concentration from 0 to 1%. The O2 flow
must not decrease more than 1 l/min through the full range.
•
If the O2 flow decreases more than 1 l/min, install a different
vaporizer and repeat the steps 1 through 4.
•
If the O2 flow decreases less than 1 l/min when testing a
different vaporizer, the malfunction is in the vaporizer that
failed the test.
•
If the O2 flow decreases more than 1 l/min with the different
vaporizer, the malfunction is most likely in the system. Do not
use the system.
5. Repeat steps 1 through 4 for each vaporizer.
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...