10 Specifications and Theory of Operation
M1132382
10-11
Gas scavenging
Pressure flow data (APL valve completely open)
Flow (l/min)
Flow (l/s)
APL pressure cmH2O
3
0.05
0.78
10
0.17
1.14
30
0.51
1.43
60
1.0
2.61
70
1.17
3.21
All scavenging
Positive pressure relief
10 cmH2O
Passive scavenging
Negative pressure relief
0.3 cmH2O
Outlet connector
30 mm male taper ISO
Active scavenging
Disposal system type
Outlet connector*
Hospital waste gas disposal
system requirements
Adjustable flow, high
vacuum
DISS EVAC
305 mmHg (12 inHg) minimum at
36 l/min flow
High flow, low vacuum
BSI 30 mm threaded
(BS6834)
50 to 80 l/min flow
Low flow, high vacuum
DISS EVAC
305 mmHg (12 inHg) minimum at
36 l/min flow
Low flow, low vacuum
12.7 mm barb
36 l/min flow
Low flow, low vacuum
25 mm barb
40 to 50 l/min flow
Low flow, low vacuum
30 mm ISO taper
male
40 to 50 l/min flow
*Other market-specific connectors may be available.
Particle filter at the outlet has a pore size of 225 microns. All flow data uses a new
filter.
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...