6 Alarms and Troubleshooting
M1132382
6-9
Minimum system
shutdown and
monitoring alarms
There are two special alarm types:
•
Minimum system monitoring alarms stop mechanical ventilation.
•
Minimum system shutdown alarms stop mechanical ventilation
and monitoring.
The software goes to minimum system monitoring when a non-
recoverable error occurs during bootup or normal operation. The
ventilator shows data and mechanical ventilation stops. ‘Monitoring
only’ appears in the Alarm message area and a specific failure
message appears in the User message area.
The following are minimum system monitoring alarms:
•
Manifold pressure sensor failure.
•
Pressure limit switch failure.
•
Flow valve (DAC) failure.
•
Flow valve (current) failure.
•
Valve power failure.
A severe malfunction causes minimum system shutdown alarms.
This condition prevents mechanical ventilation and monitoring. If this
condition occurs:
•
Ventilate manually.
•
Use a stand-alone monitor.
•
Cycle system power (On-Standby-On).
If the alarm clears, restart mechanical ventilation. If the alarm does
not clear, contact a Datex-Ohmeda
trained service representative.
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...