8 Setup and Connections
M1132382
8-3
w
To protect the patient when electrosurgical equipment is
used:
•
Monitor the correct operation of all life support and
monitoring equipment.
•
Keep backup manual ventilation available in case the
electrosurgical equipment prevents safe use of the
ventilator.
w
Do not use antistatic or electrically-conductive breathing
tubes or masks. They can cause burns if used near high
frequency surgical equipment.
w
Use only reservoir bags that comply with EN1820 on this
system.
w
Use only breathing tubes that comply with EN12342 on
this system.
w
A malfunction of the medical gas central supply system
may cause all connected devices to stop.