Aespire View
8-2
M1132382
WARNING
Datex-Ohmeda strongly recommends the use of O2
monitoring with this equipment. Refer to local standards
for mandatory monitoring.
w
European, international, and national standards require
the following monitoring be used with this system:
•
Exhaled volume monitoring.
•
O2 monitoring.
•
CO2 monitoring.
•
Anesthetic agent monitoring be used when anesthetic
vaporizers are in use.
w
Always make sure that the pipeline supply hoses and the
breathing circuit components are not toxic and will not:
•
Cause an allergic reaction in the patient.
•
React with the anesthetic gases or agent to produce
dangerous by-products.
w
To prevent incorrect values or equipment malfunction,
use only cables, hoses, and tubing approved by
Datex-Ohmeda.
w
This system operates correctly at the electrical
interference levels of IEC 60601-1-2. Higher levels can
cause nuisance alarms that may stop mechanical
ventilation.
w
To help prevent false alarms from devices with high-
intensity electrical fields:
•
Keep the electrosurgical leads away from the
breathing system, flow sensors, and O2 cell.
•
Do not allow the electrosurgical leads to contact any
part of the anesthesia system.
•
Do not use cell phones near the anesthesia system.
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...