10 Specifications and Theory of Operation
M1132382
10-13
Suction regulators (optional)
Auxiliary O2 flowmeter (optional)
Venturi Suction Regulator
Performance Category
Pharyngeal Suction
Supply
Air or O2 from system gas supply
Drive Gas Consumption*
28 l/min with pipeline drive gas at 280 kPa (41 psi)
52 l/min with pipeline drive gas at 600 kPa (87 psi)
Maximum Vacuum**
600 mmHg with pipeline drive gas at 280 kPa (41 psi)
550 mmHg with pipeline drive gas at 600 kPa (87 psi)
Maximum Flow*
29 l/min with pipeline drive gas at 280 kPa (41 psi)
32 l/min with pipeline drive gas at 600 kPa (87 psi)
Vacuum Gauge Accuracy
±5% of full scale
*Values are approximate.
**Approximate values are at sea level. Performance is reduced at higher elevations.
Continuous Suction Regulator
Performance Category
Pharyngeal Suction
Supply
External vacuum
Maximum Vacuum*
540 mmHg with external vacuum applied of
540 mmHg and 40 l/min free flow
Maximum Flow*
39 l/min with external vacuum applied of 540 mmHg
and 40 l/min free flow
Vacuum Gauge Accuracy
±5% of full scale
*Values are approximate.
Supply
O2 from system gas supply
Flow rate
0 to 10 l/min
Accuracy
±5% of full scale; not pressure compensated
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...