Aespire View
5-16
M1132382
12. Set the ventilator controls and alarm limits to clinically appropriate
levels.
13. Prepare the system.
•
Turn all vaporizers off.
•
Set the APL valve to MIN.
•
Set the Bag/Vent switch to Bag.
•
Set all flow controls to minimum.
•
Set sufficient patient suction.
•
Make sure that the breathing system is correctly connected
and not damaged.
WARNING
Flush the system with 5 l/min of O2 for at least one minute
to remove any gas mixtures or by-products from the
system.