3 Operation
M1132382
3-15
Connecting active
adjustable AGSS
The active adjustable AGSS option flow rate is limited to 30 l/min with
this option.
To use the optional active AGSS installed on the system which uses
the three-liter bag as a visual indicator, connect it as follows:
1. Connect a disposal hose to the DISS connector on the needle
valve on the bottom of the AGSS (beneath). The hose should be
flexible and reinforced to help prevent kinking and crushing.
2. Attach the other end of the hose to the hospital disposal system.
3. Attach the three-liter bag to the 30-mm auxiliary 1 port on the
bottom of the AGSS.
4. Use the needle valve to adjust the flow rate to match the amount
of gas being scavenged. Use the visual indicator bag when
adjusting the flow rate. The bag should remain partially inflated.
5. Complete the tests in the “
” section of this
manual.
Total flow sensing (optional)
The total flow sensing (TFS) option electronically measures the fresh
gas flow at the mechanical flow tubes. The total fresh gas flow
measurement displays on the bottom right corner of the screen. The
measurement specifies:
•
Total gas flow.
•
O2 gas flow.
•
Air gas flow.
•
N2O gas flow.
AB
.75
p
.1
10
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...