7 User Maintenance
M1132382
7-3
User maintenance
Datex-Ohmeda
approved service
This is the minimum level of maintenance recommended by
Datex-Ohmeda. Local regulations may contain additional
maintenance requirements. Datex-Ohmeda advocates compliance
with local regulations which meet or exceed this minimum level of
maintenance.
Breathing system maintenance
Replace any parts that are visibly cracked, chipped, distorted, or worn
when cleaning the breathing system.
Minimum Frequency
Maintenance
Daily
•
Clean the external surfaces.
•
Perform 21% O2 calibration.
•
Check the condenser reservoir (if equipped). Drain if
needed.
•
Zero the flow sensors.
Two weeks
Drain the vaporizers and discard the agent. This is not
necessary for the Tec 6 series vaporizers.
Monthly
•
Perform 100% O2 calibration.
•
Lubricate all cylinder supply tee handle threads with
Krytox or a lubricant approved for use with 100% O2.
During cleaning and
setup
Inspect the parts for damage. Replace or repair as
necessary.
Annually
Replace the external o-rings on the vaporizer ports.
As necessary
•
Install new cylinder gaskets on cylinder yokes.
•
Empty the water reservoir and replace the absorbent in
the canister.
•
Empty the overflow trap on the optional suction
regulator (if equipped).
•
Replace the circuit O2 cell (under typical use the cell
meets specifications for 1 year).
•
Replace the disposable (plastic) flow sensors (under
typical use the non-offset flow sensors meet
specifications for a minimum of three months and the
offset flow sensors meet specifications for a minimum
of six months).
•
Replace the autoclavable (metal) flow sensors (under
typical use the sensors meet specifications for a
minimum of one year).
•
Replace the receiver filter (if equipped - active gas
scavenging only).
Minimum Frequency
Maintenance
12 months
Have a Datex-Ohmeda trained service representative
complete the scheduled service maintenance checks, tests,
calibrations, and parts replacement as defined in the
Technical Reference manual.
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...