1-5
K-MINC-1000 Instructions for Use
English
3. Installation and set-up
An installation and set-up checklist has been included at the end of this section (see §4). This may be used to help
ensure correct preparation.
3.1 Unpacking
Please find the following items supplied:
1.
User manual
2.
MINC Mini Incubator
3.
Patient identification plates
(8 plates)
4.
Disposable Humidification Flask (K-MINC-CTS-S)
5.
Medical grade mains power cord
6.
External alarm connector
(on rear of device)
7.
Braided Connecting Hose
(3 metre length)
(K-MINC-BCT-10-300)
8.
USB A-B cable
(2 metre length)
9.
K-MINC-1000 Logger Software CD
Check the MINC and all items immediately upon receipt to make sure the contents are complete and that nothing
is damaged. The manufacturer will only honour claims for compensation which are forwarded immediately to the
sales representative or the authorised service agent.
Remove all items from plastic covers except the Disposable Humidification Flask which needs to be handled under
sterile conditions (refer to § 3.8).
3.2 You need to supply
• A clean source of mains power to supply the MINC. Cook Medical strongly recommends the use of a UPS with
power conditioning capabilities suitable for your local power supply system.
• Premixed medical grade gas cylinders. All gas concentrations should be within ±0.2% and cylinders should be
supplied with a certificate of analysis, refer to § 3.7.1.
• A single stage high purity gas regulator capable of supplying the gas listed above at 150 kPa (1.5 bar, 22 psi) to
the MINC inlet, refer to § 3.7.2.
• Automatic gas cylinder change over unit. Cook Medical strongly recommends the use of automatic change over
units to ensure continuity of gas supply.
• A source of sterile distilled water.
• A pressure manometer (to measure the device input gas supply pressure). This item is not required if the regulator
has gas output display gauges.
• A 9/16” spanner.
• If a longer gas connecting hose is required, contact your Cook Medical representative.
• A computer suitable to install and run the K-MINC-1000 Logger Software on.
• A USB hub for connecting multiple (up to a maximum of 10) MINC’s to a single computer.
IMPORTANT NOTE:
To reduce
the risk of damage to the MINC, the
use of an Uninterruptible Power Supply
(UPS) with power conditioning
capabilities is strongly recommended.
Furthermore, to ensure uninterrupted
device functionality, the use of
the following is also strongly
recommended:
• An automatic gas cylinder changeover
unit that can maintain a minimum
pressure of 135 kPa.
• A remote alarm to alert staff to any
failures in the gas or power supply.
IMPORTANT NOTE:
It is
important to retain packaging for
future use. (Refer to §6.5 - Return
Procedure)
IMPORTANT NOTE:
This device
has been certified by an independent
testing authority for use with USB and
Remote Alarm Cables of less than
3 metres in length.
IMPORTANT NOTE:
Use of cables
other than those provided by the
manufacturer of this equipment could
result in increased electromagnetic
emissions or decreased electromagnetic
immunity of this equipment and result
in improper operation.
WARNING:
If connecting
the MINC Mini Incubator to the
External Electrical System via the
USB cable, ensure that the External
Electrical System complies with the IEC
60601-1 safety standard or equivalent.
IMPORTANT NOTE:
Items
required but not supplied are listed
here.
IMPORTANT NOTE:
A non-
powered USB hub is suitable to connect
multiple devices.
2
3
1
8
9
7
4
5
6
1