Guidance and manufacturer’s declaration – electromagnetic emissions
The MINC Mini Incubator is intended for use in the electromagnetic environment specified below. The customer or the end user of the MINC Mini Incubator
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF emissions
CISPR 11
Group 1
The MINC Mini Incubator uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The MINC Mini Incubator is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Proximity fields from RF wireless communication equipment
Test frequency
(MHz)
Band
a
(MHz)
Service
a
Modulation
b
Maximum
power
(W)
Distance
(m)
Immunity test
level
(V/m)
Minimum
separation distance
(m)
385
360-390
TETRA 400
Pulse
modulation
b
18 Hz
1.8
0.3
27
0.3
450
430-470
GMRS 460
FRS 460
FM
C
±5 kHz deviation
1 kHz sine
2
0.3
28
0.3
710
704-787
LTE Band 13,17
Pulse
modulation
b
217 Hz
0.2
0.3
9
0.3
745
780
810
800-960
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
Pulse
modulation
b
217 Hz
2
0.3
28
0.3
870
930
1720
1700-1990
GSM 1800
CDMA 1900
DECT
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b
217 Hz
2
0.3
28
0.3
1845
1970
2450
2400-2570
Bluetooth
WLAN
802.11 b/g/n
RFID 2450
LTE Band 7
Pulse
modulation
b
217 Hz
2
0.3
28
0.3
5240
5100-5800
WLAN 802.11
a/n
Pulse
modulation
b
217 Hz
0.2
0.3
9
0.3
5500
5785
Note:
if necessary to achieve the IMMUNITY test level, the distance between the transiting antenna and the ME equipment or ME system may be reduced to
1m. The 1m test distance is permitted by IEC 61000-4-3.
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50% duty cycle square wave signal.
c