1-22
K-MINC-1000 Instructions for Use
English
7. Disposables
6.3.3 External alarm contacts
To test the external alarm contacts, ensure at least one heater chamber is active and then disconnect the MINC
from mains power. The contacts should close approximately 2 minutes later. Once the contacts close reconnect the
device to mains power and allow to operate for 150 minutes.
Open an active chamber lid and mist the chamber with cold water. During this period the temperature display
should display “Lid” and a ‘beep’ should be emitted from the MINC approximately every 30 seconds. The contacts
should close approximately 2 minutes later. If the contacts fail to close then contact your Cook representative.
6.4 Inspection by an authorised service agent
Inspections at least once a year
For ongoing operational safety of the MINC, an authorised
service agent must maintain the device annually as per Service Manual
SMA30002. The service agent will asses the operational functionality of
the gas and temperature control systems and external alarm contacts.
Authorised service agents
All services such as alterations, repairs, calibrations etc., may only be
performed by the manufacturer or by service agents who are authorised
by the manufacturer as per Service Manual SMA30002.
Liability
The manufacturer is free from all liability for the operational safety of
the MINC if the device has been wilfully opened and unauthorised
persons have performed repairs or alterations on it during the warranty
period.
Certification
The MINC owner will receive a signed certificate from the service agent
for all inspections or repairs. This certificate states the type and scope
of the services rendered, the service date and the name of the service
company.
Technical documentation
If the manufacturer provides technical documentation, this does not
authorise the user to perform repairs, adjustments or alterations to the
MINC or disposables.
6.5 Return procedure
All devices or disposables that are returned must be prepared as described below for the protection of the service
agent and for safety during transportation.
1. Clean as detailed in §6.2.
2. Seal in a plastic bag and seal within a second plastic bag.
3. Place in the original packaging.
4. Enclose the following information:
• Owner’s name
• Owner’s address
• Model type
• Serial number of the equipment (see identification plate)
• Description of the damage or fault.
The manufacturer has the right to refuse to carry out repairs if the products it receives are contaminated.
WARNING:
No user
serviceable parts inside!
WARNING: BIOLOGICAL
HAZARD.
The returned product
must be clearly marked with a
contamination warning and should
be sealed in a plastic bag and sealed
within a second plastic bag.
When shipping the MINC ensure
that the humidification flask and all
chamber contents are removed prior to
transport.
IMPORTANT NOTE:
When
returning goods, use the original
packaging. The manufacturer does not
take responsibility for damage that has
occurred during transportation if the
damage was caused by inadequate
transport packaging.
IMPORTANT NOTE:
For optimal
functioning of the incubator, use only
original disposables
Order No.
Description
K-MINC-CTS-S
Disposable Humidification Flask, gas tubing and filter set supplied.
1