K-MINC-1000 Instructions for Use
English
5.6.3 Logging data
Logging of the data to a CSV file begins as soon as a MINC is detected. The logging function cannot be turned off. A
separate data file is written for each connected device and these files are created in the data file folder. To view the
current folder, choose “Select Data File Folder” from the File menu.
From here, a new data file folder may be selected or created.
The files are named with the MINC’s name followed by its serial number. The data in these files is appended so that
they may contain data from several different runs. To clear the data, select “Clear Logged Data” from the File menu.
Data is cleared only from the actively displayed MINC.
If data files are to be kept for individual cycles, it is recommended to copy the specific MINC data file to an archiving
area and rename the copied file prior to clearing logged data files.
The copied file will contain all logged data from start-up, or the last time the file was cleared, to the time the file
was copied. If the MINC is not disconnected, and the logged data is cleared, subsequent data will continue to be
written to the cleared file.
The files may also be deleted or renamed with the Windows File Manager.
By default, measurements are logged to file at an interval of 5 minutes. Each line of the file is time and date stamped.
Alarms and other events are logged immediately. The logging interval may be set from 1 to 10 minutes by selecting
“Set Logging Interval” from the File menu.