1-23
K-MINC-1000 Instructions for Use
English
1-23
8. Technical data
Classification according to IEC 60601-1
Type of protection against electric shock:
Class I equipment
Degree of protection against harmful
ingress of solids and water:
IP31
General specifications
Power Supply:
100 - 240 VAC
Frequency:
50 - 60 Hz
Maximum power consumption:
140 VA
Alarm contact rating:
2 A / 30 VDC
Environmental operating conditions:
+18°C to +32°C
5% to 85% RH
700 hPa to 1060 hPa
Storage and transport directions:
+5°C to +40°C
10% to 75% RH
Manufactured and tested to the following standards:
IEC 60601-1: 1988 + A1: 1991 + A2: 1995
IEC 60601-1: 2005 + A1: 2012(E)
IEC60601-1-2:2014
IEC 61010-1: 2001, IEC 61010-2: 010: 2003
UL 61010-1 2ED
CSA C22.2 No. 61010-1 2ED
CSA C22.2 No. 61010-2-010 2ED
Dimensions:
Door closed: 405 mm wide x 190 mm high x 265 mm deep
Door opened: 405 mm wide x 400 mm high x 315 mm deep
Weight:
11 kg (24.3 lb)
Gas supply type:
Blend of 6% CO
2
, 5% O
2
, 89% N
2
(Cook Culture System at sea-level) or high
purity 6% CO
2
in air (recommended tolerances ±0.2%)
Gas supply pressure:
150 kPa ± 15 kPa (21.8 psi ± 2.2 psi) (1500 mbar ± 150 mbar)
Gas flow rate capability :
15 mL/min to 25 mL/min per chamber in 5 mL/min increments
Purge at 175 mL/min per chamber for 3 minutes
Gas flow rate accuracy:
±15% of flow per chamber (normal flow)
±18 mL/min per chamber (purge)
Chamber temperature capability:
35.0°C to 40.0°C in 0.1°C increments in an ambient temperature range of
+20°C to +28°C. At set point of 37°C, the ambient temperature range is
extended to +18°C to +32°C.
Chamber temperature accuracy:
±0.2°C at calibration point
Computer requirements for K-MINC-1000 Mini Incubator Logger Software
Operating system
Windows Vista, Windows XP, Windows NT, Windows 7, Windows 8 or Windows
2000
(1)
RAM
256 Mb minimum
Free hard disc space
10 Gb minimum
Processor
Pentium 4, 1.6 GHz or better
Display
1024 x 768, 256 colour or better
I/O ports:
USB port, 1.1 or 2.0 compliant
K-MINC-1000 Mini Incubator logger software specifications
Maximum connected MINC’s
10
Logging interval range
Adjustable from 1 to 10 minutes
Maximum logging time period
Only limited by free disc space
Log file format
Comma-Separated Values (CSV)
Graph Update Interval
1 second
Maximum displayed time period on screen
3 days
Temperature resolution
0.1 deg. C
Gas flow resolution
5 mL/min
Note (1): Not recommended
1