18
Function
Selection
Result
(00)
INSTANT REPORTING
(01) - (99)
DELAY
Factory Default is (
10
),
10 SECONDS
.
<NOTE>
This command is used if a fall is
detected.
If an emergency alarm is triggered by
pressing the Active Button on Fall
Sensor, the guard time delay time is
determined by
Command 42
, instead
of
Command 66
Command 67
Setting the Help Arrived Timer
This command is used to set the
countdown timer for
SUPER 32-DT35A
to
resend the alarm report if no help has
arrived after the initial report is made
(meaning the reset button has not been
pressed before the timer expires).
Function Selection
Result
(01)-(15)
1-15 MINUTES
Factory default is (
15
),
15 MINUTES
.
<NOTE>
The Help Arrived function (Command
33
) must be set as
ON
for this function
to work.
Command 75
Help Not Arrived Call
If help has not arrived after sending the
alarm report for 4 times,
SUPER 32-DT35A
will
dial
the
telephone
number
programmed by Command
75
and open
two-way communication when the call is
answered.
Enter a telephone number followed by (#)
to program the telephone number.
<NOTE>
When you program the phone
number, entering (
) will cause a 3-
second pause is for a switchboard
system, extension, etc.
The telephone number can contain up
to 20 digits, including (
).
The Help Arrived function (Command
33) must be set as ON for this function
to work.
SUPER 32-DT35A only dials this
telephone number once. If this phone
call fails, SUPER 32-DT35A will not try
again.
Command 76
Call Progress Tone
This command is used to turn on/off the
Call Progress Tones while reporting. Call
Progress Tones include all tones during
reporting such as dial tone, DTMF key
press tone, ring tone, handshake, etc.
Function
Selection
Result
(0)
OFF
(Disable)
(1)
ON
<NOTE>
On: All Call Progress Tones are
audible.
Off: Call Progress Tones are disabled
and inaudible.
This function will be disabled if
Command 35
is set as
Enable
(
Silent Mode
).
This function will be disabled if alarm
activation is made from a
Silent