FI002K0018v1530hUK – THS/21 Instruction manual for installation, use and maintenance
4
USE OF THE DEVICE
ORIGINAL
INSTRUCTION 159
4.4.8 Removing and emptying the reject bin
The reject bin must be emptied at regular intervals.
During this operation exercise extreme caution. The reject bin could be heavy. You should define a procedure for this
operation. The manufacturer declines all liability for injury to persons or damage to property caused by emptying the bin
in an incorrect manner.
The bin has a maximum bearable load, shown on a label clearly placed on the bin itself. The
operator is responsible for emptying the bin periodically before this weight is exceeded.
Example of label
If the reject bin is fitted with an electric lock, you need to unlock it using the
Unlock bin
command in Programming. You
must then open the bin within the preset
Unlock bin time
(default value 10 s, max .value 30 s) and empty it.
At the end of the operation, close the reject bin back.
4.4.8.1 THS-FBB,
THS-MBB, THS-FBV and THS-FBBV
Press the RED
Belt Stop button, if the
belt is moving;
Using the key provided, open the door of
the reject bin. Empty the bin.
Close the door again correctly.
Press the GREEN
Belt Start button to
restart the transport.
4.4.8.2 THS-MBR
Press the RED
Belt Stop button, if the
belt is moving;
Using the key provided, unlock the reject
bin to release it. Remove the bin from its
seat and empty it.
At the end of the operation, place the bin
in the guides and lock it again, with its
key.
Reset the Emergency by pressing the
BLUE
button and press the GREEN
Belt Start button to restart the
transport.
The reject bin is fitted with a safety magnetic sensor that, when the bin is removed, set the machine into the Emergency
status. After replacing the bin, to reset the system press the
Emergency Reset button.
.