3
FI002K0018v1530hUK – THS/21 Instruction manual for installation, use and maintenance
INSTALLAZIONE
110 ORIGINAL
INSTRUCTION
3.7 Description of electrical connections
3.7.1 00211ALM
card
Connector Pin Label
IN/OUT Type
Range/Max.
value
Function
J01
1
1
C
Output
Common
30 V AC/DC 2.5A
Auxiliary relay
2
NC N.C.
contact
3
NO N.O.
contact
4
C
Output
common
Test request
5
NC N.C.
contact
6
NO N.O.
contact
7
C
Output
common
Upstream conveyor relay
8
NC N.C.
contact
9
NO N.O.
contact
J02
1
C
Output
common
30 V AC/DC 2.5A
Ready relay
2
NC N.C.
contact
3
NO N.O.
contact
4
C
Output
common
Ejection relay
5
NC N.C.
contact
6
NO N.O.
contact
7
C
Output
common
Alarm relay
8
NC N.C.
contact
9
NO N.O.
contact
10
GND -
ground
-
J03
1
24 Vac
Input
24 Vac
24 V power supply input from AC/DC converter
2
+24 Vdc/ac
+24 Vdc/ac
3
0 Vdc
0 Vdc
4
- -
-
-
Reserved
5
- -
-
-
J04
Only used with the inverter module
on Conveyor Control System
J05
Reserved
J06
Reserved
J07
1
1
- -
-
-
Reserved
2
- -
-
-
3
- -
-
-
4
]
Emergency reset input
Keep jumpered
on Control Power Box
5
6
- -
-
-
Reserved
7
- -
-
-
J08
Reserved
1
Connector present only on
card, full version