FI002K0018v1530hUK – THS/21 Instruction manual for installation, use and maintenance
3
INSTALLAZIONE
ORIGINAL
INSTRUCTION 129
3.9.9.5 Automatic ejection with photocell synchronization (EM=S)
EM
Typical product
Throughput
Ejector Type
Ejection
Notes
S
Packed product
High throughput
Air jet, Pusher
Contaminated product
automatically ejected
Synchronization
photocell required
Product properly spaced
and aligned in a single line
Operation with ejection cycle activation, synchronized by photocell, for the automatic ejection of the
single contaminated pack.
In CEIA Integrated Systems fitted with an ejector, this mode determines the automatic ejection and
removal of the contaminated material in the reject bin. For this systems it is recommended to set a
transit speed lower than 60 m/min.
Programming menu
Remote
parameter
Description
Setting
Ejection
> Ejection mode
EM
Automatic removal of contaminated material
with contaminated pack synchronization
S
Ejection
> Ej. distance
ED
Distance of the ejector from the synchronization
photocell
0 - 6000 mm
Ejection
> Ejection time
ET
Ejection relay activation time
0.01 ÷ 120.00 s
Photocell at the entrance to the Probe
Generic example
S: Metal Detector Probe
N: Conveyor belt
E: Ejector
F1: Synchronization photocell
F2: Ejection check sensor
PL: Pack length
DW: Detector width
ED: Ejection distance
PD: Photocell distance
Dp: distance between packs
THS-RB systems
The ejection signal is sent directly to the ejector. In this case all conveyor sections must have the
same speed and their start must be synchronized. Parameters values can be set as indicated in
paragraphs describing other conveyor belt types.
Example with THS-RB
S:
Metal Detector Probe
E:
Ejector
F1: Synchronization photocell
F2: Ejection check sensor
PL: Pack length
DW: Detector width
ED: Ejection distance
PD: Photocell distance
Dp: distance between packs
S
E
DW
PL
Dp
PD
ED
ECPD
C
F1
F2