FI002K0018v1530hUK – THS/21 Instruction manual for installation, use and maintenance
3
INSTALLAZIONE
ORIGINAL
INSTRUCTION 125
3.9.9.1 Ejection with manual reset (EM=B)
EM
Typical product
Throughput
Ejector Type
Ejection
Notes
B
Large/heavy product
Low throughput
Transport stop
Contaminated product
to be removed
manually
-
Bulk material
Free-falling product
Operation with ejection activation (i.e. halting of the transport system) until the manual elimination of
the contaminated material and manual reset by the operator.
On CEIA Integrated Systems this mode determines the stop of the conveyor belt.
Programming menu
Remote
parameter
Description
Setting
Ejection
> Ejection mode
EM
Halting of the transport system and manual
elimination of the contaminated material
B
Ejection
> Ej. distance
ED
Distance of the contaminated pack stop area
from the centre of the Probe
0 - 6000 mm
Generic example
S: Metal Detector Probe
N: Conveyor belt
C: Contaminated pack
DW: Detector width
ED: Ejector distance
After an alarm, the Metal Detector does not check if there are any other metallic fragments in the material immediately
following.
This means that you must remove all the material between the Probe entrance and the removal area
.
To minimize waste of material it is advisable to select a low setting for the ED parameter.
In free-falling applications, or whenever the immediate activation of the ejection after an alarm is required, set ED=0.
After an alarm, if the conveyor belt is stopped, pressing STOP or one of the transport system emergency buttons
before the contaminated material has reached the predefined stop position means the Metal Detector will remain in
alarm status and cannot be reset until the belt is restarted, allowing the contaminated pack to reach the predefined stop
position.