Appendix F. Calibration and Zeroing
The currently measured value for the sensor will be displayed in the next
screen. Now place the sensor into the first known condition, and enter that
known value into the
First calibrated value
box. Press
Set First Value
. Wait
for the calibration process to measure the first value. The word
Calibrating
will be visible in the
Current Value
box until that process is complete. Now
place the sensor into the second known condition, and then enter the
corresponding known value into the
Second calibrated value
box. Press
Set
Second Value
. The calibration process measures the second point value. At
that point the datalogger calculates the new multiplier and offset and applies
them within the running program. These values are also written to the
calibration file.
F-10
Содержание LoggerNet
Страница 2: ......
Страница 30: ...Preface What s New in LoggerNet 4 xxvi...
Страница 32: ...Section 1 System Requirements 1 2...
Страница 44: ...Section 2 Installation Operation and Backup Procedures 2 12...
Страница 136: ...Section 4 Setting up Datalogger Networks 4 80...
Страница 227: ...Section 7 Creating and Editing Datalogger Programs 7 9...
Страница 298: ...Section 7 Creating and Editing Datalogger Programs 7 80...
Страница 402: ...Section 9 Automating Tasks with Task Master 9 12...
Страница 406: ...Section 9 Automating Tasks with Task Master 9 16...
Страница 450: ...Section 11 Utilities Installed with LoggerNet Admin and LoggerNet Remote 11 22...
Страница 454: ...Section 12 Optional Client Applications Available for LoggerNet 12 4...
Страница 462: ...Section 13 Implementing Advanced Communications Links 13 8...
Страница 482: ...Section 14 Troubleshooting Guide 14 20...
Страница 570: ...Appendix F Calibration and Zeroing F 16...
Страница 578: ...Appendix G Importing Files into Excel G 8...
Страница 579: ......