Appendix F. Calibration and Zeroing
The following values of the mode variable give the status of the calibration:
0 No calibration has been performed since program start
2 Calibration in progress OR first stage of two-point calibration in progress
3 Waiting for second stage of two-point calibration to begin
5 Second stage of two-point calibration in progress
6 Calibration
complete
–1 Calibration
error
–2 Measurement range error
–3 Error with the value used as the
Reps
argument in the
FieldCal
instruction.
The following values of the mode variable are used to initiate a calibration
process:
1 Start the calibration, OR start the first point of a two point calibration
4 Start the second point of a two point calibration
For Zeroing and Offset calibration, a mode value of 4 is never
used. The entire calibration process is initiated with the mode
value being set to 1.
NOTE
By properly changing the known value variables and the mode variables in a
calibrating program, a manual calibration can be performed on a sensor. Steps
for doing this are given below.
F.4.2 Using the Mode Variable for Manual Single-Point
Calibration
These steps demonstrate how to perform manual single-point calibrations
(Zeroing or Offset calibrations):
1.
Ensure the status (value of the mode variable) is 0 or 6 before you start.
a.
A number greater than 0 that is not 6 indicates that a calibration is in
progress or that the last calibration did not complete properly.
b.
A number less than 0 indicates that the calibration process
encountered an error. Resolve the error before proceeding with the
calibration then set the mode value to zero for a fresh start.
2.
Place the sensor into the zeroing or offset condition.
3.
Indicate the known offset value (if applicable) by changing the “known
value” variable to that value.
4.
Set the mode variable to 1 to initiate the calibration process.
F-6
Содержание LoggerNet
Страница 2: ......
Страница 30: ...Preface What s New in LoggerNet 4 xxvi...
Страница 32: ...Section 1 System Requirements 1 2...
Страница 44: ...Section 2 Installation Operation and Backup Procedures 2 12...
Страница 136: ...Section 4 Setting up Datalogger Networks 4 80...
Страница 227: ...Section 7 Creating and Editing Datalogger Programs 7 9...
Страница 298: ...Section 7 Creating and Editing Datalogger Programs 7 80...
Страница 402: ...Section 9 Automating Tasks with Task Master 9 12...
Страница 406: ...Section 9 Automating Tasks with Task Master 9 16...
Страница 450: ...Section 11 Utilities Installed with LoggerNet Admin and LoggerNet Remote 11 22...
Страница 454: ...Section 12 Optional Client Applications Available for LoggerNet 12 4...
Страница 462: ...Section 13 Implementing Advanced Communications Links 13 8...
Страница 482: ...Section 14 Troubleshooting Guide 14 20...
Страница 570: ...Appendix F Calibration and Zeroing F 16...
Страница 578: ...Appendix G Importing Files into Excel G 8...
Страница 579: ......