Measurement Specifications
Product Specifications
A - 12
TMS60 Operator’s Manual
Response time to heart rate change
Meets the requirements of IEC60601-2-27: Clause 201.7.9.2.9.101 b)
5).
From 80 bpm to 120 bpm: less than 11 s
From 80 bpm to 40 bpm: less than 11 s
Response to irregular rhythm
In compliance with the requirements in Clause 201.7.9.2.9.101 b) 4)
of IEC60601-2-27, the heart rate after 20 seconds of stabilization is
displayed as follows:
Ventricular bigeminy (3a): 80 ± 1 bpm
Slow alternating ventricular bigeminy (3b): 60 ± 1 bpm
Rapid alternating ventricular bigeminy (3c): 120 ± 1 bpm
Bidirectional systoles (3d): 90 ± 2 bpm
Time to alarm for tachycardia
Meets the requirements of IEC60601-2-27: Clause 201.7.9.2.9.101 b)
6).
Waveform
B1h - range: < 11 s
B1- range: < 11 s
B1d - range: < 11 s
B2h - range: < 11 s
B2- range: < 11 s
B2d- range: < 11 s
ST Segment Analysis
Measurement range
-2.0 mV to +2.0 mV RTI
Resolution
0.01 mV
Accuracy
-0.8 mV to +0.8 mV: ± 0.02 mV or ± 10%, whichever is greater
Beyond this range: Not specified
Arrhythmia Analysis
Arrhythmia analysis classifications
Asystole, V-Fib/V-Tach, V-Tach, Vent Brady, Extreme Tachy, Extreme
Brady, PVCs/min, Pauses/min, R on T, Run PVCs, Couplet, Multiform
PVC, PVC, Bigeminy, Trigeminy, Tachy, Brady, Pacer Not Pacing, Pacer
Not Capture, Missed Beats, Nonsus V-Tach, Vent Rhythm, Pause, Irr
Rhythm, A-Fib
QT Analysis
QTc formula
Bazett, Fridericia, Framingham, and Hodges
QT measurement range
200 ms to 800 ms
QT accuracy
200 ms to 800 ms: ± 30 ms
Beyond this range: Not specified
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...