12.2.1.1
Factory Default Parameter Alarm Settings
The following table lists the factory default alarm settings for all parameters.
WARNING
•
When monitoring patients that are not continuously attended by a clini-
cal operator, properly configure the alarm system and adjust alarm set-
tings as per the patient's condition.
•
Setting alarm limits to extreme values may cause the alarm system to
become ineffective. For example, high oxygen levels may predispose a
premature infant to retrolental fibroplasia. If this is a consideration do
NOT set the high alarm limit to 100%, which is equivalent to switching
the alarm off.
•
If you switch off all arrhythmia alarms, the CMS cannot give any arrhyth-
mia alarms. Always keep the patient under close surveillance.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...